Numérisation laser 3D pour machines-outils : Guide de dépannage / FAQ
Trouvez des réponses aux questions les plus fréquentes sur le dépannage des scanners laser pour machines-outils
Contact us
The team are ready to help.
-
The CNC control displays that the touchprobe is not ready after changing the scanner into the spindle
- Are you able to close this message and go on?
- If yes: Contact support if you frequently experience this problem
- If no: Check the optical status indicator of the receiver (left)
- LED red: The receiver does not receive an activation code from the CNC - contact support
- LED blinking green: Replace the batteries of the scanner
- Error persists? Contact support.
- Are you able to close this message and go on?
-
HxGN NC Measure displays that there is a LAN communication error when I want to start a scan
- Check the optical indicator for the LAN status of the receiver (right)
- LED not flashing: Check the LAN connection between PC and receiver
- LED green: Return the scanner to the machine tool magazine and call up the spindle/activate the m-function again. Wait 12s. Then try to scan again
- Check batteries
- Error persists? Contact support.
- Check the optical indicator for the LAN status of the receiver (right)
-
HxGN NC Measure displays that the scan does not contain enough data when calibrating
- Is the work offset for the calibration sphere correct?
- If yes: Clean the scanner glass and calibration sphere according to chapter 2.7.1 and 2.72 of the scanner operating instructions
- Check environment for intensive light or reflections
- Error persists? Contact support.
- Is the work offset for the calibration sphere correct?
-
HxGN NC Measure displays that the angle I want to scan is not calibrated
- Perform the sensor calibration for the tilt angles and spindle position you want to scan on your workpiece.
-
The optical glass of my scanner is soiled
- Clean the glass with a microfiber cloth carefully. If it is still soiled, follow chapter 2.7.1 of the scanner operating instructions.
-
The battery compartment is difficult to open
- Grease the sealing above the battery compartment.
-
The glass of the scanner is broken and I need a replacement
- Please contact our customer support.
-
I forgot to close the sleeve after replacing the battery and oil / liquid got inside the probe
- Remove the batteries and wipe the battery case dry
- Check whether the scanner is still functioning
- Contact customer support if the scanner is not working.
-
I am not sure if my scanner is still accurate
- Please contact customer support to perform an accuracy check. We recommend checking the scanner’s performance and accuracy on a regular basis.
-
How accurate is my scan?
- Please see scanner operating instructions.
-
It is not possible to send a program from HxGN NC Measure to the CNC
- General: Check ethernet connection between the control and PC
- FANUC:
- Set control to EDIT mode
- Heidenhain:
- Set control to MANUAL mode
- Check DNC connection tool for active connection to CNC
- The scan program must not be opened
- SIEMENS
- A scan program which is currently running is not allowed.